Categories
Recent Posts
OM-A EECP Machine Operation Manual
1.Wrap cuffs:
1)Let patient lie on treatment bed
2)Wrap the cuffs around calves,lower thighs and upper thighs which with more muscle.The tighter you wrapped,the better treatment effect.
3)Attach the electrodes to skin like following picture shows:
4)Place the finger sensor to a middle finger
2.Turn on machine:
1)Turn on power switch at the back side of machine
2)Click”treatment” to begin EECP treatment.
3.Treatment:
1)Click”standby”:valves begin to work
2)Click”start” and it will appear a dialog,Click”new” or “browse” to input patient’s data(“New” for new patients,”browse” for the patients data that saved in the computer before) and set treatment time,once click”ok”,bed will working now.
3)Adjust the pressure according to patient’s systolic pressure,The initial pressure recommend is the systolic pressure plus 50mmHg.
4.Finished Treatment:
1)The bed will automatically stop working when treatment time is expired.
2)Take out ECG Cable and SPO2 cable
3)Take off cuffs
5.Turn off machine
1)Click”stop” to stop EECP treatment then click”Exit” will back to the main interface
2)Click”Exit” will turn off the screen
3)Turn off the power switch at the back side of machine.
For more details of EECP machine,please contact me through sales@eecpcn.com or WhatsApp/Mobile:+86-18520667125